Table 3 Comparison of ELP and certified lipid values for the NIST SRM 1951c reference material

From: The extended lipid panel assay: a clinically-deployed high-throughput nuclear magnetic resonance method for the simultaneous measurement of lipids and Apolipoprotein B

 

Total Cholesterol

(mg/dL)

Triglycerides

(mg/dL)

HDL Cholesterol

(mg/dL)

LDL Cholesterol

(mg/dL)

Level

ELP

valuea

Ref.

valueb

Biasc

%

ELP

value

Ref.

value

Bias

%

ELP

value

Ref.

value

Bias

%

ELP

value

Ref.

value

Bias

%

1

157.3

(2.9)

152.4

(1.8)

3.2

153.0

(1.7)

152.0

(3.2)

0.7

43.3

(1.2)

41.0

(0.9)

5.6

87.0

(1.7)

86.4

(1.4)

0.7

2

243.0

(2.6)

241.4

(2.8)

0.7

139.7

(1.5)

145.4

(3.2)

−3.9

66.3

(1.2)

64.9

(1.7)

2.2

152.0 (1.0)

143.8

(2.1)

5.7

  1. aELP values are means (standard deviation) of triplicate measurement. bReference values are means (95% confidence intervals). cBias is percent difference between ELP and reference value